Loading (50 kb)...'
(continued)
(v) Allow for transport delays and the slowest instrument's full response.
(vi) Repeat the steps in paragraphs (d)(3)(i) through (v) of this section to record seven full cycles, ending with zero gas flowing to the analyzers.
(vii) Stop recording.
(e) Performance evaluation. (1) If you chose to demonstrate compliance with paragraph (c)(1) of this section, use the data from paragraph (d)(3) of this section to calculate the mean rise time, T10–90, and mean fall time, T90–10, for each of the analyzers. Multiply these times (in seconds) by their respective recording frequencies in Hertz (1/second). The value for each result must be at least 5. If the value is less than 5, increase the recording frequency or adjust the flows or design of the sampling system to increase the rise time and fall time as needed. You may also configure digital filters to increase rise and fall times.
(2) If a measurement system fails the criterion in paragraph (e)(1) of this section, ensure that signals from the system are updated and recorded at a frequency of at least 5 Hz.
(3) If a measurement system fails the criteria in paragraphs (e)(1) and (2) of this section, you may use the continuous analyzer system only if the deficiency does not adversely affect your ability to show compliance with the applicable standards.
§ 1065.309 Continuous gas analyzer uniform response verification.
top
(a) Scope and frequency. If you use more than one continuous gas analyzer to quantify a gaseous constituent, you must perform this verification. For example, if you determine NMHC as the difference between continuous THC and CH4 measurements, you must perform this verification on your NMHC measurement system. As another example if you determine NOX as the sum of separate continuous measurements of NO and NO2, you must perform this verification on your NOX measurement system. Also, you must perform this verification if you use one continuous analyzer to apply an interference compensation algorithm to another continuous gas analyzer. Perform this verification after initial installation or major maintenance. Also perform this verification if you reconfigure your system in a way that would change system response. For example, perform this verification if you add a significant volume to the transfer lines by increasing their length or by adding a filter; or if you change the frequency at which you sample and record gas-analyzer concentrations.
(b) Measurement principles. This procedure verifies the time-alignment and uniform response of combined continuous gas measurements.
(c) System requirements. Demonstrate that combined continuous concentration measurements have a uniform rise and fall during a simultaneous to a step change in both concentrations. During a system response to a rapid change in multiple gas concentrations, demonstrate that the t50 times of all combined analyzers all occur at the same recorded second of data or between the same two recorded seconds of data.
(d) Procedure. Use the following procedure to verify the response of a continuous gas analyzer system:
(1) Instrument setup. Follow the analyzer system manufacturer's start-up and operating instructions. Adjust the system as needed to optimize performance.
(2) Equipment setup. Using minimal gas transfer line lengths between all connections, connect a zero-air source to the inlet of a 100 °C heated line. Connect the heated line outlet to one inlet of a 100 °C heated fast-acting 3-way valve (2 inlets, 1 outlet). Using a gas divider, equally blend an NO–CO–CO2–C3H8–CH4 (balance N2) span gas with a span gas of NO2 (balance N2). Connect the gas divider outlet to the inlet of a 50 °C heated line. Connect the heated line outlet to the inlet of a 50 °C gas bubbler filled with distilled water. Connect the bubbler outlet to another heated line at 100 °C. Connect the outlet of the 100 °C line to the other inlet of the 3-way valve. Connect the valve outlet to an overflow at the gas analyzer system's probe or to an overflow fitting between the probe and transfer line to all the analyzers being verified.
(3) Data collection. (i) Switch the valve to flow zero gas.
(ii) Allow for stabilization, accounting for transport delays and the slowest instrument's full response.
(iii) Start recording data at the frequency used during emission testing.
(iv) Switch the valve to flow span gas.
(v) Allow for transport delays and the slowest instrument's full response.
(vi) Repeat the steps in paragraphs (d)(3)(i) through (v) of this section to record seven full cycles, ending with zero gas flowing to the analyzers.
(vii) Stop recording.
(e) Performance evaluations. Perform the following evaluations:
(1) Uniform response evaluation. (i) Calculate the mean rise time, t10–90, mean fall time, t90–10 for each analyzer.
(ii) Determine the maximum mean rise and fall times for the slowest responding analyzer in each combination of continuous analyzer signals that you use to determine a single emission concentration.
(iii) If the maximum rise time or fall time is greater than one second, verify that all other gas analyzers combined with it have mean rise and fall times of at least 75% of that analyzer's response.
(iv) If any analyzer has shorter rise or fall times, disperse that signal so that it better matches the rise and fall times of the slowest signal with which it is combined. We recommend that you perform dispersion using SAE 2001–01–3536 (incorporated by reference in §1065.1010) as a guide.
(v) Repeat this verification after optimizing your systems to ensure that you dispersed signals correctly. If after repeated attempts at dispersing signals your system still fails this verification, you may use the continuous analyzer system if the deficiency does not adversely affect your ability to show compliance with the applicable standards.
(2) Time alignment evaluation. (i) After all signals are adjusted to meet the uniform response evaluation, determine the second at which—or the two seconds between which—each analyzer crossed the midpoint of its response, t50.
(ii) Verify that all combined gas analyzer signals are time-aligned such that all of their t50 times occurred at the same second or between the same two seconds in the recorded data.
(iii) If your system fails to meet this criterion, you may change the time alignment of your system and retest the system completely. If after changing the time alignment of your system, some of the t50 times still are not aligned, take corrective action by dispersing analyzer signals that have the shortest rise and fall times.
(iv) If some t50 times are still not aligned after repeated attempts at dispersion and time alignment, you may use the continuous analyzer system if the deficiency does not adversely affect your ability to show compliance with the applicable standards.
Measurement of Engine Parameters and Ambient Conditions
top
§ 1065.310 Torque calibration.
top
(a) Scope and frequency. Calibrate all torque-measurement systems including dynamometer torque measurement transducers and systems upon initial installation and after major maintenance. Use good engineering judgment to repeat the calibration. Follow the torque transducer manufacturer's instructions for linearizing your torque sensor's output. We recommend that you calibrate the torque-measurement system with a reference force and a lever arm.
(b) Recommended procedure. (1) Reference force quantification. Use either a set of dead-weights or a reference meter such as strain gage or a proving ring to quantify the reference force, NIST-traceable within ±0.5% uncertainty.
(2) Lever-arm length quantification. Quantify the lever arm length, NIST-traceable within ±0.5% uncertainty. The lever arm's length must be measured from the centerline of the dynamometer to the point at which the reference force is measured. The lever arm must be perpendicular to gravity (i.e., horizontal), and it must be perpendicular to the dynamometer's rotational axis. Balance the lever arm's torque or quantify its net hanging torque, NIST-traceable within ±1% uncertainty, and account for it as part of the reference torque.
(c) Dead-weight calibration. This technique applies a known force by hanging known weights at a known distance along a lever arm. Make sure the weights' lever arm is perpendicular to gravity (i.e., horizontal) and perpendicular to the dynamometer's rotational axis. Apply at least six calibration-weight combinations for each applicable torque-measuring range, spacing the weight quantities about equally over the range. Oscillate or rotate the dynamometer during calibration to reduce frictional static hysteresis. Determine each weight's force by multiplying its NIST-traceable mass by the local acceleration of Earth's gravity (using this equation: force = mass · acceleration). The local acceleration of gravity, ag, at your latitude, longitude, and elevation may be determined by entering position and elevation data into the U.S. National Oceanographic and Atmospheric Administration's surface gravity prediction Web site at http://www.ngs.noaa.gov/cgi-bin/grav_pdx.prl. If this Web site is unavailable, you may use the equation in §1065.630, which returns the local acceleration of gravity based on a given latitude. In this case, calculate the reference torque as the weights' reference force multiplied by the lever arm reference length (using this equation: torque = force · lever arm length).
(d) Strain gage or proving ring calibration. This technique applies force either by hanging weights on a lever arm (these weights and their lever arm length are not used) or by operating the dynamometer at different torques. Apply at least six force combinations for each applicable torque-measuring range, spacing the force quantities about equally over the range. Oscillate or rotate the dynamometer during calibration to reduce frictional static hysteresis. In this case, the reference torque is determined by multiplying the reference meter force output by its effective lever-arm length, which you measure from the point where the force measurement is made to the dynamometer's rotational axis. Make sure you measure this length perpendicular to gravity (i.e., horizontal) and perpendicular to the dynamometer's rotational axis.
§ 1065.315 Pressure, temperature, and dewpoint calibration.
top
(a) Calibrate instruments for measuring pressure, temperature, and dewpoint upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration, as follows:
(1) Pressure. We recommend temperature-compensated, digital-pneumatic, or deadweight pressure calibrators, with data-logging capabilities to minimize transcription errors. We recommend using calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(2) Temperature. We recommend digital dry-block or stirred-liquid temperature calibrators, with datalogging capabilities to minimize transcription errors. We recommend using calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(3) Dewpoint. We recommend a minimum of three different temperature-equilibrated and temperature-monitored calibration salt solutions in containers that seal completely around the dewpoint sensor. We recommend using calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(b) You may remove system components for off-site calibration. We recommend specifying calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
Flow-Related Measurements
top
§ 1065.320 Fuel-flow calibration.
top
(a) Calibrate fuel-flow meters upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration.
(b) You may also develop a procedure based on a chemical balance of carbon or oxygen in engine exhaust.
(c) You may remove system components for off-site calibration. When installing a flow meter with an off-site calibration, we recommend that you consider the effects of the tubing configuration upstream and downstream of the flow meter. We recommend specifying calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
§ 1065.325 Intake-flow calibration.
top
(a) Calibrate intake-air flow meters upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration. We recommend using a calibration subsonic venturi, ultrasonic flow meter or laminar flow element. We recommend using calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(b) You may remove system components for off-site calibration. When installing a flow meter with an off-site calibration, we recommend that you consider the effects of the tubing configuration upstream and downstream of the flow meter. We recommend specifying calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(c) If you use a subsonic venturi or ultrasonic flow meter for intake flow measurement, we recommend that you calibrate it as described in §1065.340.
§ 1065.330 Exhaust-flow calibration.
top
(a) Calibrate exhaust-flow meters upon initial installation. Follow the instrument manufacturer's instructions and use good engineering judgment to repeat the calibration. We recommend that you use a calibration subsonic venturi or ultrasonic flow meter and simulate exhaust temperatures by incorporating a heat exchanger between the calibration meter and the exhaust-flow meter. If you can demonstrate that the flow meter to be calibrated is insensitive to exhaust temperatures, you may use other reference meters such as laminar flow elements, which are not commonly designed to withstand typical raw exhaust temperatures. We recommend using calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(b) You may remove system components for off-site calibration. When installing a flow meter with an off-site calibration, we recommend that you consider the effects of the tubing configuration upstream and downstream of the flow meter. We recommend specifying calibration reference quantities that are NIST-traceable within 0.5% uncertainty.
(c) If you use a subsonic venturi or ultrasonic flow meter for raw exhaust flow measurement, we recommend that you calibrate it as described in §1065.340.
§ 1065.340 Diluted exhaust flow (CVS) calibration.
top
(a) Overview. This section describes how to calibrate flow meters for diluted exhaust constant-volume sampling (CVS) systems.
(b) Scope and frequency. Perform this calibration while the flow meter is installed in its permanent position. Perform this calibration after you change any part of the flow configuration upstream or downstream of the flow meter that may affect the flow-meter calibration. Perform this calibration upon initial CVS installation and whenever corrective action does not resolve a failure to meet the diluted exhaust flow verification (i.e., propane check) in §1065.341.
(c) Reference flow meter. Calibrate a CVS flow meter using a reference flow meter such as a subsonic venturi flow meter, a long-radius ASME/NIST flow nozzle, a smooth approach orifice, a laminar flow element, a set of critical flow venturis, or an ultrasonic flow meter. Use a reference flow meter that reports quantities that are NIST-traceable within ±1% uncertainty. Use this reference flow meter's response to flow as the reference value for CVS flow-meter calibration.
(d) Configuration. Do not use an upstream screen or other restriction that could affect the flow ahead of the reference flow meter, unless the flow meter has been calibrated with such a restriction.
(e) PDP calibration. Calibrate a positive-displacement pump (PDP) to determine a flow-versus-PDP speed equation that accounts for flow leakage across sealing surfaces in the PDP as a function of PDP inlet pressure. Determine unique equation coefficients for each speed at which you operate the PDP. Calibrate a PDP flow meter as follows:
(1) Connect the system as shown in Figure 1 of this section.
(2) Leaks between the calibration flow meter and the PDP must be less than 0.3% of the total flow at the lowest calibrated flow point; for example, at the highest restriction and lowest PDP-speed point.
(3) While the PDP operates, maintain a constant temperature at the PDP inlet within ±2% of the mean absolute inlet temperature, T in.
(4) Set the PDP speed to the first speed point at which you intend to calibrate.
(5) Set the variable restrictor to its wide-open position.
(6) Operate the PDP for at least 3 min to stabilize the system. Continue operating the PDP and record the mean values of at least 30 seconds of sampled data of each of the following quantities:
(i) The mean flow rate of the reference flow meter, n ref. This may include several measurements of different quantities, such as reference meter pressures and temperatures, for calculating n ref.
(ii) The mean temperature at the PDP inlet, T in.
(iii) The mean static absolute pressure at the PDP inlet, p in.
(iv) The mean static absolute pressure at the PDP outlet, p out.
(v) The mean PDP speed, f nPDP.
(7) Incrementally close the restrictor valve to decrease the absolute pressure at the inlet to the PDP, p in.
(8) Repeat the steps in paragraphs (e)(6) and (7) of this section to record data at a minimum of six restrictor positions reflecting the full range of possible in-use pressures at the PDP inlet.
(9) Calibrate the PDP by using the collected data and the equations in §1065.640.
(10) Repeat the steps in paragraphs (e)(6) through (9) of this section for each speed at which you operate the PDP.
(11) Use the equations in §1065.642 to determine the PDP flow equation for emission testing.
(12) Verify the calibration by performing a CVS verification (i.e., propane check) as described in §1065.341.
(13) Do not use the PDP below the lowest inlet pressure tested during calibration.
(f) CFV calibration. Calibrate a critical-flow venturi (CFV) to verify its discharge coefficient, Cd, at the lowest expected static differential pressure between the CFV inlet and outlet. Calibrate a CFV flow meter as follows:
(1) Connect the system as shown in Figure 1 of this section.
(2) Start the blower downstream of the CFV.
(3) While the CFV operates, maintain a constant temperature at the CFV inlet within ±2% of the mean absolute inlet temperature, T in.
(4) Leaks between the calibration flow meter and the CFV must be less than 0.3 % of the total flow at the highest restriction.
(5) Set the variable restrictor to its wide-open position.
(6) Operate the CFV for at least 3 min to stabilize the system. Continue operating the CFV and record the mean values of at least 30 seconds of sampled data of each of the following quantities:
(i) The mean flow rate of the reference flow meter, n ref. This may include several measurements of different quantities, such as reference meter pressures and temperatures, for calculating n ref.
(ii) Optionally, the mean dewpoint of the calibration air, T dew. See §1065.640 for permissible assumptions.
(iii) The mean temperature at the venturi inlet, T in.
(iv) The mean static absolute pressure at the venturi inlet, p in.
(v) The mean static differential pressure between the CFV inlet and the CFV outlet, ?p CFV.
(7) Incrementally close the restrictor valve to decrease the absolute pressure at the inlet to the CFV, pin.
(8) Repeat the steps in paragraphs (f)(6) and (7) of this section to record mean data at a minimum of ten restrictor positions, such that you test the fullest practical range of ?p CFV expected during testing. We do not require that you remove calibration components or CVS components to calibrate at the lowest possible restrictions.
(9) Determine Cd and the lowest allowable ?p CFV as described in §1065.640.
(10) Use Cd to determine CFV flow during an emission test. Do not use the CFV below the lowest allowed ?p CFV, as determined in §1065.640.
(11) Verify the calibration by performing a CVS verification (i.e., propane check) as described in §1065.341.
(12) If your CVS is configured to operate more than one CFV at a time in parallel, calibrate your CVS by one of the following:
(i) Calibrate every combination of CFVs according to this section and §1065.640. Refer to §1065.642 for instructions on calculating flow rates for this option.
(ii) Calibrate each CFV according to this section and §1065.640. Refer to §1065.642 for instructions on calculating flow rates for this option.
(g) SSV calibration. Calibrate a subsonic venturi (SSV) to determine its calibration coefficient, Cd , for the expected range of inlet pressures. Calibrate an SSV flow meter as follows:
(1) Connect the system as shown in Figure 1 of this section.
(2) Start the blower downstream of the SSV.
(3) Leaks between the calibration flow meter and the SSV must be less than 0.3 % of the total flow at the highest restriction.
(4) While the SSV operates, maintain a constant temperature at the SSV inlet within ±2 % of the mean absolute inlet temperature, T in.
(5) Set the variable restrictor or variable-speed blower to a flow rate greater than the greatest flow rate expected during testing. You may not extrapolate flow rates beyond calibrated values, so we recommend that you make sure the Reynolds number, Re#, at the SSV throat at the greatest calibrated flow rate is greater than the maximum Re# expected during testing.
(6) Operate the SSV for at least 3 min to stabilize the system. Continue operating the SSV and record the mean of at least 30 seconds of sampled data of each of the following quantities:
(i) The mean flow rate of the reference flow meter, n ref. This may include several measurements of different quantities, such as reference meter pressures and temperatures, for caculating n ref.
(ii) Optionally, the mean dewpoint of the calibration air, T dew. See §1065.640 for permissible assumptions.
(iii) The mean temperature at the venturi inlet, T in .
(iv) The mean static absolute pressure at the venturi inlet, p in.
(v) Static differential pressure between the static pressure at the venturi inlet and the static pressure at the venturi throat, ?p SSV.
(7) Incrementally close the restrictor valve or decrease the blower speed to decrease the flow rate.
(8) Repeat the steps in paragraphs (g)(6) and (7) of this section to record data at a minimum of ten flow rates.
(9) Determine a functional form of Cd versus Re# by using the collected data and the equations in §1065.640.
(10) Verify the calibration by performing a CVS verification (i.e., propane check) as described in §1065.341 using the new Cd versus Re# equation.
(11) Use the SSV only between the minimum and maximum calibrated flow rates.
(12) Use the equations in §1065.642 to determine SSV flow during a test.
(h) Ultrasonic flow meter calibration. [Reserved]
View or download PDF
§ 1065.341 CVS and batch sampler verification (propane check).
top
(a) A propane check serves as a CVS verification to determine if there is a discrepancy in measured values of diluted exhaust flow. A propane check also serves as a batch-sampler verification to determine if there is a discrepancy in a batch sampling system that extracts a sample from a CVS, as described in paragraph (g) of this section. Using good engineering judgment and safe practices, this check may be performed using a gas other than propane, such as CO2 or CO. A failed propane check might indicate one or more problems that may require corrective action, as follows:
(1) Incorrect analyzer calibration. Re-calibrate, repair, or replace the FID analyzer.
(2) Leaks. Inspect CVS tunnel, connections, fasteners, and HC sampling system, and repair or replace components.
(3) Poor mixing. Perform the verification as described in this section while traversing a sampling probe across the tunnel's diameter, vertically and horizontally. If the analyzer response indicates any deviation exceeding ±2% of the mean measured concentration, consider operating the CVS at a higher flow rate or installing a mixing plate or orifice to improve mixing.
(4) Hydrocarbon contamination in the sample system. Perform the hydrocarbon-contamination verification as described in §1065.520.
(5) Change in CVS calibration. Perform an in-situ calibration of the CVS flow meter as described in §1065.340.
(6) Other problems with the CVS or sampling verification hardware or software.
Inspect the CVS system, CVS verification hardware, and software for discrepancies.
(b) A propane check uses either a reference mass or a reference flow rate of C3H8 as a tracer gas in a CVS. Note that if you use a reference flow rate, account for any non-ideal gas behavior of C3H8 in the reference flow meter. Refer to §1065.640 and §1065.642, which describe how to calibrate and use certain flow meters. Do not use any ideal gas assumptions in §1065.640 and §1065.642. The propane check compares the calculated mass of injected C3H8 using HC measurements and CVS flow rate measurements with the reference value.
(c) Prepare for the propane check as follows:
(1) If you use a reference mass of C3H8 instead of a reference flow rate, obtain a cylinder charged with C3H8. Determine the reference cylinder's mass of C3H8 within ±0.5% of the amount of C3H8 that you expect to use.
(2) Select appropriate flow rates for the CVS and C3H8.
(3) Select a C3H8 injection port in the CVS. Select the port location to be as close as practical to the location where you introduce engine exhaust into the CVS. Connect the C3H8 cylinder to the injection system.
(4) Operate and stabilize the CVS.
(5) Preheat or precool any heat exchangers in the sampling system.
(6) Allow heated and cooled components such as sample lines, filters, chillers, and pumps to stabilize at operating temperature.
(7) You may purge the HC sampling system during stabilization.
(8) If applicable, perform a vacuum side leak verification of the HC sampling system as described in §1065.345.
(9) You may also conduct any other calibrations or verifications on equipment or analyzers.
(d) Zero, span, and verify contamination of the HC sampling system, as follows:
(1) Select the lowest HC analyzer range that can measure the C3H8 concentration expected for the CVS and C3H8 flow rates.
(2) Zero the HC analyzer using zero air introduced at the analyzer port.
(3) Span the HC analyzer using C3H8 span gas introduced at the analyzer port.
(4) Overflow zero air at the HC probe or into a fitting between the HC probe and the transfer line.
(5) Measure the stable HC concentration of the HC sampling system as overflow zero air flows. For batch HC measurement, fill the batch container (such as a bag) and measure the HC overflow concentration.
(6) If the overflow HC concentration exceeds 2 µmol/mol, do not proceed until contamination is eliminated. Determine the source of the contamination and take corrective action, such as cleaning the system or replacing contaminated portions.
(7) When the overflow HC concentration does not exceed 2 µmol/mol, record this value as xHCpre and use it to correct for HC contamination as described in §1065.660.
(e) Perform the propane check as follows:
(1) For batch HC sampling, connect clean storage media, such as evacuated bags.
(2) Operate HC measurement instruments according to the instrument manufacturer's instructions.
(3) If you will correct for dilution air background concentrations of HC, measure and record background HC in the dilution air.
(4) Zero any integrating devices.
(5) Begin sampling, and start any flow integrators.
(6) Release the contents of the C3H8 reference cylinder at the rate you selected. If you use a reference flow rate of C3H8, start integrating this flow rate.
(7) Continue to release the cylinder's contents until at least enough C3H8 has been released to ensure accurate quantification of the reference C3H8 and the measured C3H8.
(8) Shut off the C3H8 reference cylinder and continue sampling until you have accounted for time delays due to sample transport and analyzer response.
(9) Stop sampling and stop any integrators.
(f) Perform post-test procedure as follows:
(1) If you used batch sampling, analyze batch samples as soon as practical.
(2) After analyzing HC, correct for contamination and background.
(3) Calculate total C3H8 mass based on your CVS and HC data as described in §1065.650 and §1065.660, using the molar mass of C3H8, MC3H8, instead the effective molar mass of HC, MHC.
(4) If you use a reference mass, determine the cylinder's propane mass within ±0.5% and determine the C3H8 reference mass by subtracting the empty cylinder propane mass from the full cylinder propane mass.
(5) Subtract the reference C3H8 mass from the calculated mass. If this difference is within ±2.0 % of the reference mass, the CVS passes this verification. If not, take corrective action as described in paragraph (a) of this section.
(g) Batch sampler verification. You may repeat the propane check to verify a batch sampler, such as a PM secondary dilution system.
(1) Configure the HC sampling system to extract a sample near the location of the batch sampler's storage media (such as a PM filter). If the absolute pressure at this location is too low to extract an HC sample, you may sample HC from the batch sampler pump's exhaust. Use caution when sampling from pump exhaust because an otherwise acceptable pump leak downstream of a batch sampler flow meter will cause a false failure of the propane check.
(2) Repeat the propane check described in this section, but sample HC from the batch sampler.
(3) Calculate C3H8 mass, taking into account any secondary dilution from the batch sampler.
(4) Subtract the reference C3H8 mass from the calculated mass. If this difference is within ±5% of the reference mass, the batch sampler passes this verification. If not, take corrective action as described in paragraph (a) of this section.
§ 1065.345 Vacuum-side leak verification.
top
(a) Scope and frequency. Upon initial sampling system installation, after major maintenance, and before each test according to subpart F of this part for laboratory tests and according to subpart J of this part for field tests, verify that there are no significant vacuum-side leaks using one of the leak tests described in this section.
(b) Measurement principles. A leak may be detected either by measuring a small amount of flow when there should be zero flow, or by detecting the dilution of a known concentration of span gas when it flows through the vacuum side of a sampling system.
(c) Low-flow leak test. Test a sampling system for low-flow leaks as follows:
(1) Seal the probe end of the system by taking one of the following steps:
(i) Cap or plug the end of the sample probe.
(ii) Disconnect the transfer line at the probe and cap or plug the transfer line.
(iii) Close a leak-tight valve in-line between a probe and transfer line.
(2) Operate all vacuum pumps. After stabilizing, verify that the flow through the vacuum-side of the sampling system is less than 0.5 % of the system's normal in-use flow rate. You may estimate typical analyzer and bypass flows as an approximation of the system's normal in-use flow rate.
(d) Dilution-of-span-gas leak test. Test any analyzer, other than a FID, for dilution of span gas as follows, noting that this configuration requires an overflow span gas system:
(1) Prepare a gas analyzer as you would for emission testing.
(2) Supply span gas to the analyzer port and verify that it measures the span gas concentration within its expected measurement accuracy and repeatability.
(3) Route overflow span gas to one of the following locations in the sampling system:
(i) The end of the sample probe.
(ii) Disconnect the transfer line at the probe connection, and overflow the span gas at the open end of the transfer line.
(iii) A three-way valve installed in-line between a probe and its transfer line, such as a system overflow zero and span port.
(4) Verify that the measured overflow span gas concentration is within the measurement accuracy and repeatability of the analyzer. A measured value lower than expected indicates a leak, but a value higher than expected may indicate a problem with the span gas or the analyzer itself. A measured value higher than expected does not indicate a leak.
CO and CO2 Measurements
top
§ 1065.350 H2O interference verification for CO2 NDIR analyzers.
top
(a) Scope and frequency. If you measure CO2 using an NDIR analyzer, verify the amount of H2O interference after initial analyzer installation and after major maintenance.
(b) Measurement principles. H2O can interfere with an NDIR analyzer's response to CO2.
If the NDIR analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, simultaneously conduct these other measurements to test the compensation algorithms during the analyzer interference verification.
(c) System requirements. A CO2 NDIR analyzer must have an H2O interference that is within ±2% of the flow-weighted mean CO2 concentration expected at the standard, though we strongly recommend a lower interference that is within ±1%.
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the CO2 NDIR analyzer as you would before an emission test.
(2) Create a water-saturated test gas by bubbling zero air that meets the specifications in §1065.750 through distilled water in a sealed vessel at (25 ±10) °C.
(3) Introduce the water-saturated test gas upstream of any sample dryer, if one is used during testing.
(4) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(5) While the analyzer measures the sample's concentration, record 30 seconds of sampled data. Calculate the arithmetic mean of this data. The analyzer meets the interference verification if this value is within ±2% of the flow-weighted mean concentration of CO2 expected at the standard.
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your CO2 sampling system and your emission-calculation procedures, the H2O interference for your CO2 NDIR analyzer always affects your brake-specific emission results within ±0.5% of each of the applicable standards.
(2) You may use a CO2 NDIR analyzer that you determine does not meet this verification, as long as you try to correct the problem and the measurement deficiency does not adversely affect your ability to show that engines comply with all applicable emission standards.
§ 1065.355 H2O and CO2 interference verification for CO NDIR analyzers.
top
(a) Scope and frequency. If you measure CO using an NDIR analyzer, verify the amount of H2O and CO2 interference after initial analyzer installation and after major maintenance.
(b) Measurement principles. H2O and CO2 can positively interfere with an NDIR analyzer by causing a response similar to CO. If the NDIR analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, simultaneously conduct these other measurements to test the compensation algorithms during the analyzer interference verification.
(c) System requirements. A CO NDIR analyzer must have combined H2O and CO2 interference that is within ±2 % of the flow-weighted mean concentration of CO expected at the standard, though we strongly recommend a lower interference that is within ±1%.
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the CO NDIR analyzer as you would before an emission test.
(2) Create a water-saturated CO2 test gas by bubbling a CO2 span gas through distilled water in a sealed vessel at (25 ±10) °C.
(3) Introduce the water-saturated CO2 test gas upstream of any sample dryer, if one is used during testing.
(4) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(5) While the analyzer measures the sample's concentration, record its output for 30 seconds. Calculate the arithmetic mean of this data.
(6) Multiply this mean value by the ratio of expected CO2 to span gas CO2 concentration. In other words, estimate the flow-weighted mean dry concentration of CO2 expected during testing, and then divide this value by the concentration of CO2 in the span gas used for this verification. Then multiply this ratio by the mean value recorded during this verification.
(7) The analyzer meets the interference verification if the result of paragraph (d)(6) of this section is within ±2 % of the flow-weighted mean concentration of CO expected at the standard.
(e) Exceptions. The following exceptions apply:
(1) You may omit this verification if you can show by engineering analysis that for your CO sampling system and your emission calculations procedures, the combined CO2 and H2O interference for your CO NDIR analyzer always affects your brake-specific CO emission results within ±0.5 % of the applicable CO standard.
(2) You may use a CO NDIR analyzer that you determine does not meet this verification, as long as you try to correct the problem and the measurement deficiency does not adversely affect your ability to show that engines comply with all applicable emission standards.
Hydrocarbon Measurements
top
§ 1065.360 FID optimization and verification.
top
(a) Scope and frequency. For all FID analyzers perform the following steps:
(1) Calibrate a FID upon initial installation. Repeat the calibration as needed using good engineering judgment.
(2) Optimize a FID's response to various hydrocarbons after initial analyzer installation and after major maintenance.
(3) Determine a FID's methane (CH4) response factor after initial analyzer installation and after major maintenance.
(4) Verify methane (CH4) response within 185 days before testing.
(b) Calibration. Use good engineering judgment to develop a calibration procedure, such as one based on the FID-analyzer manufacturer's instructions and recommended frequency for calibrating the FID. Alternately, you may remove system components for off-site calibration. Calibrate using C3H8 calibration gases that meet the specifications of §1065.750. We recommend FID analyzer zero and span gases that contain approximately the flow-weighted mean concentration of O2 expected during testing. If you use a FID to measure methane (CH4) downstream of a nonmethane cutter, you may calibrate that FID using CH4 calibration gases with the cutter. Regardless of the calibration gas composition, calibrate on a carbon number basis of one (C1). For example, if you use a C3H8 span gas of concentration 200 µmol/mol, span the FID to respond with a value of 600 µmol/mol.
(c) FID response optimization. Use good engineering judgment for initial instrument start-up and basic operating adjustment using FID fuel and zero air. Heated FIDs must be within their required operating temperature ranges. Optimize FID response at the most common analyzer range expected during emission testing. Optimization involves adjusting flows and pressures of FID fuel, burner air, and sample to minimize response variations to various hydrocarbon species in the exhaust. Use good engineering judgment to trade off peak FID response to propane calibration gases to achieve minimal response variations to different hydrocarbon species. For an example of trading off response to propane for relative responses to other hydrocarbon species, see SAE 770141 (incorporated by reference in §1065.1010). Determine the optimum flow rates for FID fuel, burner air, and sample and record them for future reference.
(d) CH4 response factor determination. Since FID analyzers generally have a different response to CH4 versus C3H8, determine each FID analyzer's CH4 response factor, RFCH4, after FID optimization. Use the most recent RFCH4 measured according to this section in the calculations for HC determination described in §1065.660 to compensate for CH4 response. Determine RFCH4 as follows, noting that you do not determine RFCH4 for FIDs that are calibrated and spanned using CH4 with a nonmethane cutter:
(1) Select a C3H8 span gas that meets the specifications of §1065.750. Record the C3H8 concentration of the gas.
(2) Select a CH4 span gas that meets the specifications of §1065.750. Record the CH4 concentration of the gas.
(3) Start and operate the FID analyzer according to the manufacturer's instructions.
(4) Confirm that the FID analyzer has been calibrated using C3H8. Calibrate on a carbon number basis of one (C1). For example, if you use a C3H8 span gas of concentration 200 µmol/mol, span the FID to respond with a value of 600 µmol/mol.
(5) Zero the FID with a zero gas that you use for emission testing.
(6) Span the FID with the C3H8 span gas that you selected under paragraph (d)(1) of this section.
(7) Introduce at the sample port of the FID analyzer, the CH4 span gas that you selected under paragraph (d)(2) of this section.
(8) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the analyzer and to account for its response.
(9) While the analyzer measures the CH4 concentration, record 30 seconds of sampled data. Calculate the arithmetic mean of these values.
(10) Divide the mean measured concentration by the recorded span concentration of the CH4 calibration gas. The result is the FID analyzer's response factor for CH4, RFCH4.
(e) FID methane (CH4) response verification. If the value of RFCH4 from paragraph (d) of this section is within ±5.0% of its most recent previously determined value, the FID passes the methane response verification. For example, if the most recent previous value for RFCH4 was 1.05 and it changed by +0.05 to become 1.10 or it changed by -0.05 to become 1.00, either case would be acceptable because +4.8% is less than +5.0%.
(1) Verify that the pressures and flow rates of FID fuel, burner air, and sample are each within ±0.5% of their most recent previously recorded values, as described in paragraph (c) of this section. You may adjust these flow rates as necessary. Determine a new RFCH4 as described in paragraph (d) of this section.
(2) If RFCH4 is still not within ±5.0% of its most recently determined value after adjusting flow rates, re-optimize the FID response as described in paragraph (c) of this section.
(3) Determine a new RFCH4 as described in paragraph (d) of this section. Use this new value of RFCH4 in the calculations for HC determination, as described in §1065.660.
§ 1065.362 Non-stoichiometric raw exhaust FID O2 interference verification.
top
(a) Scope and frequency. If you use FID analyzers for raw exhaust measurements from engines that operate in a non-stoichiometric mode of combustion (e.g., compression-ignition, lean-burn), verify the amount of FID O2 interference upon initial installation and after major maintenance.
(b) Measurement principles. Changes in O2 concentration in raw exhaust can affect FID response by changing FID flame temperature. Optimize FID fuel, burner air, and sample flow to meet this verification. Verify FID performance with the compensation algorithms for FID O2 interference that you have active during an emission test.
(c) System requirements. Any FID analyzer used during testing must meet the FID O2 interference verification according to the procedure in this section.
(d) Procedure. Determine FID O2 interference as follows:
(1) Select two span reference gases that meet the specifications in §1065.750 and contain C3H8 near 100% of span for HC. You may use CH4 span reference gases for FIDs calibrated on CH4 with a nonmethane cutter. Select the two balance gas concentrations such that the concentrations of O2 and N2 represent the minimum and maximum O2 concentrations expected during testing.
(2) Confirm that the FID analyzer meets all the specifications of §1065.360.
(3) Start and operate the FID analyzer as you would before an emission test. Regardless of the FID burner's air source during testing, use zero air as the FID burner's air source for this verification.
(4) Zero the FID analyzer using the zero gas used during emission testing.
(5) Span the FID analyzer using the span gas used during emission testing.
(6) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of sampled data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, then proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(7) Check the analyzer response using the span gas that has the minimum concentration of O2 expected during testing. Record the mean response of 30 seconds of stabilized sample data as xO2minHC.
(8) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of stabilized sample data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, then proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(9) Check the analyzer response using the span gas that has the maximum concentration of O2 expected during testing. Record the mean response of 30 seconds of stabilized sample data as xO2maxHC.
(10) Check the zero response of the FID analyzer using the zero gas used during emission testing. If the mean zero response of 30 seconds of stabilized sample data is within ±0.5% of the span reference value used in paragraph (d)(5) of this section, then proceed to the next step; otherwise restart the procedure at paragraph (d)(4) of this section.
(11) Calculate the percent difference between xO2maxHC and its reference gas concentration. Calculate the percent difference between xO2minHC and its reference gas concentration. Determine the maximum percent difference of the two. This is the O2 interference.
(12) If the O2 interference is within ±1.5%, then the FID passes the O2 interference check; otherwise perform one or more of the following to address the deficiency:
(i) Select zero and span gases for emission testing that contain higher or lower O2 concentrations.
(ii) Adjust FID burner air, fuel, and sample flow rates. Note that if you adjust these flow rates to meet the O2 interference verification, you must re-verify with the adjusted flow rates that the FID meets the CH4 response factor verification according to §1065.360.
(iii) Repair or replace the FID.
(iv) Demonstrate that the deficiency does not adversely affect your ability to demonstrate compliance with the applicable emission standards.
§ 1065.365 Nonmethane cutter penetration fractions.
top
(a) Scope and frequency. If you use a FID analyzer and a nonmethane cutter (NMC) to measure methane (CH4), determine the nonmethane cutter's penetration fractions of methane, PFCH4, and ethane, PFC2H6. Perform this verification after installing the nonmethane cutter. Repeat this verification within 185 days of testing to verify that the catalytic activity of the cutter has not deteriorated. Note that because nonmethane cutters can deteriorate rapidly and without warning if they are operated outside of certain ranges of gas concentrations and outside of certain temperature ranges, good engineering judgment may dictate that you determine a nonmethane cutter's penetration fractions more frequently.
(b) Measurement principles. A nonmethane cutter is a heated catalyst that removes nonmethane hydrocarbons from the exhaust stream before the FID analyzer measures the remaining hydrocarbon concentration. An ideal nonmethane cutter would have PFCH4 of 1.000, and the penetration fraction for all other hydrocarbons would be 0.000, as represented by PFC2H6. The emission calculations in §1065.660 use this section's measured values of PFCH4 and PFC2H6 to account for less than ideal NMC performance.
(c) System requirements. We do not limit NMC penetration fractions to a certain range. However, we recommend that you optimize a nonmethane cutter by adjusting its temperature to achieve PFCH4 > 0.95 and PFC2H6 < 0.02 as determined by paragraphs (d) and (e) of this section, as applicable. If we use a nonmethane cutter for testing, it will meet this recommendation. If adjusting NMC temperature does not result in achieving both of these specifications simultaneously, we recommend that you replace the catalyst material.
Use the most recently determined penetration values from this section to calculate HC emissions according to §1065.660 and §1065.665 as applicable.
(d) Procedure for a FID calibrated with the NMC. If your FID arrangement is such that a FID is always calibrated to measure CH4 with the NMC, then span that FID with the NMC cutter using a CH4 span gas, set that FID's CH4 penetration fraction, PFCH4, equal to 1.0 for all emission calculations, and determine its ethane (C2H6) penetration fraction, PFC2H6. as follows:
(1) Select a CH4 gas mixture and a C2H6 analytical gas mixture and ensure that both mixtures meet the specifications of §1065.750. Select a CH4 concentration that you would use for spanning the FID during emission testing and select a C2H6 concentration that is typical of the peak NMHC concentration expected at the hydrocarbon standard or equal to THC analyzer's span value.
(2) Start, operate, and optimize the nonmethane cutter according to the manufacturer's instructions, including any temperature optimization.
(3) Confirm that the FID analyzer meets all the specificationsof §1065.360. (continued)